Título | The performance of a prototype device designed to evaluate general quality parameters of X-ray equipment |
Ano | 2014 |
Mês | Feb. |
Data de Acesso | 11 maio 2024 |
Tipo de Trabalho | journal article |
Tipo Secundário | PRE PI |
Número de Arquivos | 1 |
Tamanho | 2097 KiB |
|
2. Contextualização | |
Autor | 1 Murata, Camila H. 2 Fernandes, Diego Camilo 3 Lavínia, N. C. 4 Caldas, L. V. E. 5 Pires, S. R. 6 Medeiros, R. B. |
Grupo | 1 2 CMC-ETES-SPG-INPE-MCTI-GOV-BR |
Afiliação | 1 Coordenadoria de Física e Higiene das Radiações (CFHR), Departamento de Diagnóstico por Imagem, Universidade Federal de São Paulo 2 Instituto Nacional de Pesquisas Espaciais (INPE) 3 Mosaico Ind. e Com Eletro Eletrônico Ltda; Santo André 4 Instituto de Pesquisas Energéticas e Nucleares (IPEN) 5 Universidade Federal de São Paulo 6 Universidade Federal de São Paulo |
Endereço de e-Mail do Autor | 1 camila.murata@gmail.com 2 dcfernandes@metrosp.com.br |
Endereço de e-Mail | dcfernandes@metrosp.com.br |
Revista | Radiation Physics and Chemistry |
Volume | 95 |
Páginas | 101–105 |
Histórico (UTC) | 2013-12-12 17:17:57 :: lattes -> administrator :: 2014-02-11 12:18:39 :: administrator :: 2013 -> 2014 2018-06-04 23:39:21 :: administrator -> marcelo.pazos@inpe.br :: 2014 |
|
3. Conteúdo e estrutura | |
É a matriz ou uma cópia? | é a matriz |
Estágio do Conteúdo | concluido |
Transferível | 1 |
Tipo do Conteúdo | External Contribution |
Tipo de Versão | publisher |
Palavras-Chave | multifunction X-Ray analysis system solid state detector ionizing chamber quality control tests |
Resumo | The performance of radiological equipment can be assessed using non-invasive methods and portable instruments that can analyze an X-ray beam with just one exposure. These instruments use either an ionization chamber or a state solid detector (SSD) to evaluate X-ray beam parameters. In Brazil, no such instruments are currently being manufactured; consequently, these instruments come at a higher cost to users due to importation taxes. Additionally, quality control tests are time consuming and impose a high workload on the X-ray tubes when evaluating their performance parameters. The assessment of some parameters, such as the half-value layer (HVL), requires several exposures; however, this can be reduced by using a SSD that requires only a single exposure. One such SSD uses photodiodes designed for high X-ray sensitivity without the use of scintillation crystals. This sensitivity allows one electron-hole pair to be created per 3.63 eV of incident energy, resulting in extremely high and stable quantum efficiencies. These silicon photodiodes operate by absorbing photons and generating a flow of current that is proportional to the incident power. The aim of this study was to show the response of the solid sensor PIN RD100A detector in a multifunctional X-ray analysis system that is designed to evaluate the average peak voltage (kVp), exposure time, and HVL of radiological equipment. For this purpose, a prototype board that uses four SSDs was developed to measure kVp, exposure time, and HVL using a single exposure. The reproducibility and accuracy of the results were compared to that of different X-ray beam analysis instruments. The kVp reproducibility and accuracy results were 2% and 3%, respectively; the exposure time reproducibility and accuracy results were 2% and 1%, respectively; and the HVL accuracy was ±2%. The prototype's methodology was able to calculate these parameters with appropriate reproducibility and accuracy. Therefore, the prototype can be considered a multifunctional instrument that can appropriately evaluate the performance of radiological equipment. |
Área | ETES |
Arranjo | urlib.net > BDMCI > Fonds > Produção pgr ATUAIS > CMC > The performance of... |
Conteúdo da Pasta doc | acessar |
Conteúdo da Pasta source | não têm arquivos |
Conteúdo da Pasta agreement | não têm arquivos |
|
4. Condições de acesso e uso | |
Idioma | en |
Arquivo Alvo | 1-s2.0-S0969806X13001758-main.pdf |
Grupo de Usuários | administrator lattes marcelo.pazos@inpe.br |
Grupo de Leitores | administrator marcelo.pazos@inpe.br |
Visibilidade | shown |
Permissão de Leitura | deny from all and allow from 150.163 |
Permissão de Atualização | não transferida |
|
5. Fontes relacionadas | |
Repositório Espelho | iconet.com.br/banon/2006/11.26.21.31 |
Unidades Imediatamente Superiores | 8JMKD3MGPCW/3F2UALS |
URL (dados não confiáveis) | http://dx.doi.org/10.1016/j.radphyschem.2013.03.041 |
Divulgação | WEBSCI; PORTALCAPES; COMPENDEX; SCOPUS. |
Acervo Hospedeiro | dpi.inpe.br/plutao@80/2008/08.19.15.01 |
|
6. Notas | |
Notas | Proceedings of the 12th International Symposium on Radiation Physics (ISRP 2012) |
Campos Vazios | alternatejournal archivingpolicy archivist callnumber copyholder copyright creatorhistory descriptionlevel format isbn lineage mark nextedition number orcid parameterlist parentrepositories previousedition previouslowerunit progress project resumeid rightsholder schedulinginformation secondarydate secondarykey secondarymark session shorttitle sponsor subject tertiarymark tertiarytype |
|
7. Controle da descrição | |
e-Mail (login) | marcelo.pazos@inpe.br |
atualizar | |
|